For best results, the S-SCORT III must be reconnected to the charging source after each use and remain on charge until needed. Disconnect only for portable use.
Check the condition of the battery frequently, see the battery test on page 6 of the owners manual.
The S-SCORT III Charger with LED Battery Indicator (#80533) works exclusively with the S-SCORT III (model 74000) and plugs into a standard outlet.
The charger charges the battery only. The unit will not run off of the charger.
The LED battery indicator on the charger will show red when charging the battery and will change to green once the battery is charged. SSCOR recommends that the S-SCORT III remain on charge when not in use, even if the LED battery indicator of the charger is green. The battery will not overcharge.
One other note: The charger's default light color is green. It changes to red when plugged into the device and back to green once the battery is charged.
The suction device will run on vehicle power when the unit is connected directly to the vehicle battery via the DC power cord (# 80665). Consult the owners manual for complete information.
AC charging should bring the battery to a dependable working charge in 4 to 6 hours. The charging indicator light on the front panel of the unit indicates the S-SCORT III is receiving electrical input from the AC charger or vehicle. It does not reflect the condition of the battery. Check the charging Indicator light to be sure you have a good electrical connection. Consult the owners manual for complete information.
SSCOR Batteries
Each component in the S-SCORT III is critical for delivering effective vacuum when you need it. You can count on the S-SCORT III to give you the airway clearing suction that you need.
Each part in the S-SCORT III aspirator has been carefully chosen and tested for reliability.
Each S-SCORT III comes with a SSCOR labeled battery that has been chosen for this purpose.
When you need to replace the battery, which we recommend that you do every three years, we recommend that you replace it with a SSCOR labeled battery.
Reliable, dependable, that is our SSCOR standard.
WARNING:
This device contains a Sealed Lead Acid Battery. Sealed Lead Acid Batteries contain lead and lead compounds, chemicals known to the State of California to cause cancer and reproductive harm. Batteries also contain other chemicals known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.
SSCOR Inc. | 2125 N. Madera Rd. Unit C, Simi Valley, CA 93065 USA | T 818-504-4054 | Email: sales@sscor.com